Specifications
TruO2 OTC
Accuracy
- SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits.
- Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits.
- Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits.
- Clinical studies have shown that the device’s accuracy is not affected by differences in skin pigmentation.
Operating Conditions
- Temperatures no lower or higher than -5°C to 40°C
(23°F to 104°F). - Humidity no lower or higher than 10 to 95%.
Battery Life
- Approximately 2,200 uses or stored for 1 month.
- If stored for more than 30 days, remove the batteries.
Accuracy Summary
SaO2 Measured Accuracy
70 – 100% 2.27
70 – 80% 2.85
80 – 90% 1.86
90 – 100% 1.97
2500A VET
Oxygen Saturation Display Range
- 0% to 100% SpO2
Pulse Rate Display Range
- 18 to 450 beats per minute (BPM)
Saturation Declared Accuracy (Arms)*
- 70 – 100% ±3 digits for the Model 2000SL, 2000SA, and 2000T sensors.
- Below 70% is not specified for all sensors.
Pulse Rate Declared Accuracy
- ±3% ±3 digits
Alarm Volume
- High: 70 dBA
- Low: 55 dBA
Informational Tone Volume
- High: 65 dBA
- Low: 45 dBA
Measurement Wavelengths and Output Power**
- Red: 660 nanometers @ 0.8 mW maximum average
- Infrared: 910 nanometers @ 1.2 mW maximum average
Indicators
- Pulse Quality: LED, tricolor
- Numeric Displays: 3-digit, 7-segment LEDs, red
- Low Battery: LED, amber
- Alarm Bar: LED, bicolor
- Alarm Silence: LED, amber
Temperature
- Operating: -20 to +50 °C (-4 to +122 °F)
- Storage/Transportation: -40 to +70 °C (-40 to +158 °F)
Humidity
- Operating: 10% to 95% non-condensing
- Storage/Transportation: 10% to 95% non-condensing
Altitude
- Operating: Up to 4,000 meters (13,123 feet)
- Hyperbaric Pressure: Up to 4 atmospheres
Power Requirements
- Four 1.5V AA-size alkaline batteries (60 hours typical operation)
- NiMH rechargeable battery pack (40 hours typical operation)
Dimensions
- 13.8 cm H x 7.0 cm W x 3.2 cm D (5.4 in H x 2.8 in W x 1.3 in D)
Weight
- 213 g (7.5 oz) (with alkaline batteries) 233 g (8.2 oz) (with NiMH rechargeable battery pack)
Classifications per ANSI/AAMI ES60601-1 and CAN/CSA-C22.2 No. 60601-1
- Type of Protection: Internally powered (on battery power)
- Degree of Protection: Type BF-Applied Part
- Mode of Operation: Continuous
Enclosure Degree of Ingress Protection
- IP32
This device is not made with natural rubber latex.
* ± 1 Arms represents approximately 68% of measurements at zero bias.